4-4
Z MED Medical Air System
4107 9022 67.00
Problem
Possible Causes
Solution
Dew point degrada-
tion
Incorrect purge gas flow
Inlet gas temperature to high
Liquid entering the dryer inlet
Purge exhaust muffler restricted
Desiccant is contaminated
Piping component leaks at dryer
outlet or downstream of dryer
outlet
Check purge orifice for blockage.
Clean or replace as required
Check compressor aftercooler
Inspect pre-filter cartridges. Replace
if necessary.
Inspect Zero-loss drain valves. Re-
place if necessary.
Replace muffler
Shutdown and depressurize the
dryer. Inspect the desiccant and
replace if fouled. Inspect prefilters if
fouling is noted.
Soap bubble test the dryer outlet
and downstream piping. Repair all
leaks noted.
Backpressure on a
desiccant chamber
during regeneration
cycle
Dirty or fouled muffler
Switch off power, remove purge muf
-
fler and clean or replace.
Maximum Capacity
System undersized
Incorrect pressure setting
Faulty pressure transducer
System piping leaks
Both dryers on line
Contact
BeaconMedæs
Adjust Pressure band parameters
Replace transducer
Repair leaks
Valve off one dryer
High receiver water
level
Manual isolation valve closed
Automatic drain valve not func-
tioning
Open valve
Check electrical hookup
Service, clean or replace internal
parts
Replace valve assembly