B-7
Z MED Medical Air System
4107 9022 67.00
Name
Description
Controller
Default Value
Beacon
Medaes Value
'Setpoint 1'
95 psi*
95 psi*
‘Setpoint 2’
95 psi
95 psi
‘Indirect stop level 1’***
4 psi
4 psi
'Direct stop level 1'
15 psi
15 psi
'Indirect stop level 2'***
4 psi
4 psi
‘Direct stop level 2’
15 psi
15 psi
‘Proportional band’
10 percent
10 percent
'Integration time'
6 seconds
6 seconds
'Minimum speed'
Minimum motor speed
Varies
Default
'Maximum rpm reduction'
100 percent
100 percent
Appendix B - Compressor Controller (Mk4)
B.7.1 Compressor Motor Listing
B.7.2 Parameter Listing
Name
Description
Controller
Default Value
Beacon
Medaes Value
'Fan Motor Start/Day'
Limits the number of times the fan motor
can be started and stopped in a day
240
240
‘Minimum stop time’
Minimum time between start and stop
5 seconds
5 seconds
‘Power recovery time’
Maximum allowed power down time to
accept the Automatic Restart After Voltage
Failure (ARAVF)
3600 seconds
oo! sec**
‘Restart Delay’
Start delay after voltage is restored
5 seconds
5 seconds
‘Communication Time-out’
30 seconds
30 seconds
‘Setpoint 1’
95 psi*
95 psi*
'Setpoint 2'
95 psi
95 psi
‘Indirect stop level 1’
4 psi
4 psi
'Direct stop level 1'***
15 psi
15 psi
‘Indirect stop level 2’
4 psi
4 psi
'Direct stop level 2'***
15 psi
15 psi
‘Proportional band’
10 percent
percent
'Integration time'
6 seconds
seconds
‘Minimum speed’
Minimum motor speed
Varies
Default
‘Maximum rpm reduction’
100 percent
100 percent
* Typical settings for lead compressor. Lower these settings by 5 psi for each remaining compressor so
each has a different local set point.
** Controller display will show “oo! Sec” when “Auto Restart” parameter is set to infinite.
*** It may be necessary to use FSP (Field Service Program) to set this parameter. Contact BeaconMedaes
Technical Support for additional information.