3-2
Z MED Medical Air System
4107 9022 67.00
3.2 Initial System Start-Up
Before starting the system for the first time all
parameters in the Central Control System on the
Dryer module and individual compressor mod-
ule controllers must be verified/set-up.
• See Appendix A.6 for complete list of param-
eters for the Central Control System.
• See Appendix B.5 or Appendix C.5 for com-
plete list of parameters for the Compressor
Controllers.
3.2.1 Compressor Initial Start-Up
(See also ZT/ZR Instruction Book provided with
compressor)
1. A number of VCI (Volatile Corrosion Inhibi-
tor) foam sheets may be provided inside the
bodywork and electrical cabinet to protect
the compressor against corrosion. Remove
the plates.
2. The compressor and motor are secured to
the frame, immobilizing the vibration damp-
ers during transport (FIG 3.2.1a). After install-
ing compressor, remove the transport fixa
-
tions which are painted red. Unscrew bolt (1)
and remove transport bushing (5). Six trans-
port fixations are provided: two at the gear
casing side, two at the motor side, and two to
support the intercooler.
3.
Check that the gear casing is filled with oil,
the level should be in the middle of the sight-
glass. Add oil via the filler plug if necessary.
(FIG 3.2.1b)
4. Check that the electrical connections cor-
respond to the local codes. The installation
must be grounded and protected by fuses in
all phases. An isolating switch must be pro-
vided.
5. Check the connections at the primary sides
of transformers (T1 and T2). S
ee the Service
Diagram in Electrical Connections section of
the ZT/ZR Instruction Book.
On aircooled com-
pressors, check the settings of circuit breaker
(Q15) and overload relay (F21).
6. Close the manual drain valve of the conden-
sate traps.
7.
On “ZR” Only:
Check that the cooling water
drain valves (customer installed) in the inlet
and outlet lines are closed. Open the water
inlet and outlet valves (customer installed).
Open the water flow regulating valves (1), (2)
, and (3) and check for water flow. (FIG 3.2.1c)
8.
On “All” compressors:
Switch on the volt-
age. Start the compressor and stop it im-
mediately. Check for correct direction of ro-
3.0 System Operation
FIG 3.2.1a Transport Fixations
Gear Casing Side (1) / Motor Side (2)
Components
1 Bolt
2 Nut
3 Bolt
4 Nut
5 Bush
6
Motor support or gear casing
FIG 3.2.1b Oil System Components
Components
FC
Filter Plug
SG
Sight Glass
Dmo Oil drain valve, gear casing