D-7
Z MED Medical Air System
4107 9022 67.00
Appendix D - Desiccant Dryer
D.4.5.2 Main Screen
When working in normal operation the display
shows the following image:
Icon
Description
Rotating arrow
Active when the dryer is in its working phases
Still vertical bars
Visualized when the dryer is in the FREEZE status
Still square
Visualized when the dryer is stopped
Phase
Text displayed on screen
Translation
Stop
STOPPED
STOPPED
Shifting
SHIFTING
SHIFTING
Adsorbing
ADSORBING
ADSORBING
Pressure relief
DEPRESS
DEPRESS
Regeneration
REGENERAT
REGENERAT
Pressure equalize
EQUALIZE
EQUALIZE
Freeze
FREEZE
FREEZE
Standby
STANDBY
STANDBY
Blowing off
BLOW OFF
BLOW OFF
• In the upper left area of the screen, an
animated icon is present:
• In the upper right area, the current working
phase countdown and the measured dew
point are visualized.
- When the PDP (pressure dewpoint)
sensor is not enabled, the dewpoint is not
displayed.
- Dewpoint Transmitter is controlled at the
Central Control Panel (CCP).
• In the lower area, the current working
phase of each vessel is displayed.
The possible working phases are: