2-7
Z MED Medical Air System
4107 9022 67.00
2.0 Installation
2.8.5 Line Terminations
Correctly “terminating” the CAN network is im-
portant for a properly functioning network. This
is done by inserting so-called line termination
resistors at the beginning and at the end of the
network cable. In the connectors as specified in
FIG 2.8.5, this termination function is included
and can be set by means of a small terminator
switch. Make sure this switch is in the “ON” posi-
tion for both the first and last device, and in the
“OFF” position for all other devices. Both connec-
tor types are equipped with this switch.
FIG 2.8.4 Middle Wiring Connections
FIG 2.8.4a Connector Switch Setting