D-9
Z MED Medical Air System
4107 9022 67.00
Appendix D - Desiccant Dryer
D.4.5.5 Editing Parameters
During parameters editing (Section D.4.5.3,
menu item 2 “Settings”), the display shows an
image similar to the following:
Module Involved
Cause
Message on
the
Display
How to Clear the
Alarm
Alarm LED
Status
Service
Running hours >
programmed
service hours A, B
or C
Running hours
>
service hours X
(X
= A, B or C)
Perform the requested mainte-
nance (A, B, or C). Next, enter the
Alarm menu, select Service Alarm,
press the ENTER button and keep it
pressed during minimum
5 seconds
Blinking
24V supply
Supply voltage
< 18V
24V supply is in
fault
Contact your
supplier
Blinking
Controller
Controller
hardware
fault
Internal error
Reset all settings to the default
value. Switch off the controller and
switch it on again. Call your sup-
plier if the fault persists
Lit con-
tinuously
Possible alarm causes:
In the upper right area, the current parameter
number and the total modifiable parameters
number are visualized.
In the lower area, the current parameter de-
scription and value are displayed.
Editor button functions:
Button
Function
LEFT
• Selects the previous parameter
• Decrements the value
RIGHT
• Selects the next parameter
• Increments the value
ENTER
• Starts the parameter editing
• Selects the numerical data digits
•
Confirms the modified value
Only the unprotected parameters
can be dis-
played and modified! Example of settings that
are accessible to the end user are language and
the display time-out.
Pressing the ENTER button for at least 2 seconds
ends the editing procedure. The display will
return to the selection menu.