A-2
Z MED Medical Air System
4107 9022 67.00
A.2.3 Priority Selection
Two priority schemes are available.
BeaconMedæs
recommends setting all com-
pressors to Level 1.
A.2.4 Equalized Running Hours
Within a group of compressors that have the
same priority, the compressor with the least
running hours will be started first, compressors
with the most running hours will be stopped
first.
A.2.5 Compressor Availability
Before a compressor in a group is started, an
availability check is performed. This means that
only the compressors that are running in unloaded
condition or are ready to be started will be tak
en
into account.
A compressor can be unavailable due to:
• Minimum stop time
• Shut-down condition
• Emergency stop pressed
• No communication
A.2.6 Automatic Restart After Voltage
Failure
The controller has a function for automatic
restart after voltage failure. For controllers
leaving the factory, this function is made active.
After power restoring, the controller restarts
regulation and returns to the condition
experienced before the power failure.
Programmed delay times and simultaneous
start prevention will be taken into account.
A.3 CAN Network Connection
See “
Section 2.0 Installation, Section 2.7
” for
network cabling of compressors and cable end
connector wiring and assembly.
Appendix A - Central Control System
FIG A.4 Dryer Central Control Panel