4-2
Z MED Medical Air System
4107 9022 67.00
Problem
Possible Causes
Solution
Air temperature
above normal
Inlet temperature to high due to
bad room ventilation or recircula-
tion of cooling air
Air filter clogged
On “ZR”
, insufficient cooling water
flow
On “ZR”
,
restriction in cooling
water system due to formation of
scale or settling down of dirt
Improve ventilation of compressor
room and avoid cooling air recircula-
tion
Replace filter
Check water temperature and in-
crease cooling water flow
Contact
BeaconMedæs
Condensate is not
discharged from con-
densate traps during
operation
Discharge flex line clogged
Electronic water drain malfunc-
tioning
Check and correct as necessary
Contact
BeaconMedæs
Motor overheating
Low voltage
Defective motor
Check for proper supply voltage
Contact
BeaconMedæs
Compressor runs hot Wrong rotation
Incorrect pressure setting
Faulty compressor valves
Intake filter clogged
Switch motor leads
Adjust pressure band parameters
Contact
BeaconMedæs
Clean or replace
Wrong direction of
rotation
Motor wired incorrectly
Switch motor leads
Low discharge pres-
sure
System piping leaks
Defective pressure gauge
Faulty pressure transducer
Intake filter clogged
Repair leaks
Replace gauge
Adjust or replace
Clean or replace