B.5 Parameters – Fixed Speed
Compressor
The compressor parameters are separated into
the two following sections: B.5.1 Parameters
and B.5.2 Configuration. The two sections are
located under “Modify Parameters” on the
compressor controller. See B.4 for on-screen
navigation diagram.
B.5.1 Parameter Listing
Appendix B - Compressor Controller (Mk4)
Name
Description
Controller
Default Value
Beacon
Medaes Value
‘Pressure Band 1 Low’
Low pressure level of Pressure band 1
7 bar
90 psi
'Pressure Band 1 High'
High pressure level of Pressure band 1
8 bar
105 psi
'Pressure Band 2 Low'
Low pressure level of Pressure band 1
7 bar
90 psi
‘Pressure Band 2 High’
High pressure level of Pressure band 1
8 bar
105 psi
'Start/Load Reac. IDT
Start + Load Reaction time of connection
channel 1 Time which the compressor has
to run in load condition after a start com-
mand
20 seconds
20 seconds
'Y-Time'
10 seconds
10 seconds
‘Load Reaction’
Time which the compressor has to run in
load condition after start command
10 seconds
10 seconds
'Stop Reaction'
Time which the compressor has to stop
after a stop command
35 seconds
35 seconds
'Unload Reaction'
Time which the compressor has to run in
unload condition after an unload command
5 seconds
5 seconds
'Load Delay'
0 seconds
0 seconds
'Restart Delay'
Start delay after voltage is restored
5 seconds
5 seconds
‘Number of Starts/Day’
Limits the number of times the compressor
can be started and stopped in a day
240
240
'Minimum Stop Time'
Time which the compressor has to stop
before restarting
20 seconds
20 seconds
‘Communication Timeout’
30 seconds
30 seconds