C-8
Z MED Medical Air System
4107 9022 67.00
Appendix C - Compressor Controller (Mk5)
C.6.1 Regulation
C.6.2 Network
Name
Description
Controller
Default Value
Beacon
Medaes Value
Reduced Setpoint
102 psi¹
95 psi¹
Setpoint 1
102 psi¹
95 psi¹
Indirect Stop Level 1
116 psi¹
116 psi¹
Direct Stop Level 1
123 psi¹
123 psi¹
Setpoint 2
Indirect Stop Level 2
Direct Stop Level 2
Setpoint Used
Set Point 1
Set Point 1
Digital Pressure Band Selection
Requires password to activate
(4735)
Off
Off
Minimum Stop Time
5 seconds
5 seconds
Minimum Speed
Varies
Default
Minimum Speed Factor
0%
0%
Proportional Band
10%
10%
Integration Time
6 seconds
6 seconds
1. Typical settings shown for lead compressor. Lower these settings by 5 psi for each remaining
compressor so that all compressors have a different local set point.
Name
Description
Controller
Default Value
Beacon
Medaes Value
CAN
Ethernet
CAN
CAN¹
CAN must be off to modify
address or profile.
Off
Indefinite¹
CAN Address
1
Node ID
Communication Profile
Mk4
Mk4
1. CAN must be off to change CAN address and communication profile.