A-14
Z MED Medical Air System
4107 9022 67.00
A.6.2.5 Counters
The “Counters” menu displays the Module (sys-
tem) total run time and the run time for each
compressor in the module. (See FIG A.6.2.5
Counters Menu Flow Diagram)
A.6.2.6 Test
The “Test” menu allows the user to test the
display and test the operation of the 9 NO/
NC outputs from the Elektronikon. Refer to
the Electrical Schematic included with the unit
to identify the operations controlled by each
output from the Elektronikon. Note: The Ele-
ktronikon must be in “local” mode before you
can test the controllers’ outputs. To do so, press
“F3” when at the main screen. The display “Loc.”
just above the “F3” button will turn off and the
FIG A.6.2.4 Measured Data Menu Flow Diagram (Continued)
center display will change from “automatic
operation” to “local operation” (See FIG A.5.1
ES-6 Status for respective operation symbols).
Once in local mode navigate to the “Test” menu
to test outputs (See FIG A.6.2.6 Test menu flow
diagram). To return the system to “automatic
operation” press the start button on the left
side of the controller.
A.6.2.7 Modify Parameters - Parameters
The Parameters section is password protected
and can only be changed by a certified
BeaconMedæs
service technician. Contact
BeaconMedæs
Technical Services department
at 1-888-4MEDGAS (888-463-3427) for assis-
tance. (See FIG A.6.2.7 Parameters menu flow
diagram).
Appendix A - Central Control System