![Bard E-Luminexx Vascular Stent Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/bard/e-luminexx-vascular-stent/e-luminexx-vascular-stent_instructions-for-use-manual_3855873008.webp)
• During delivery system preparation, ensure that the red safety clip remains in place until the stent is ready to be deployed. If the red
safety clip has been removed or becomes inadvertently detached from the grip, D
Do
o N
No
oT
T use the device.
9
9..5
5
S
Seelleeccttiioon
n ooff D
Deep
pllooyym
meen
ntt M
Meetth
hood
d::
• Determine whether you will use the P
P
E
ER
RFFo
oR
R
M
MA
AX
XX
X®® G
Grriip
p for stent deployment. (See instructions for “Stent Deployment with the
P
ERFoR
MAXX® Grip”, Section 9.8.)
• If selecting “The Conventional Method” of stent deployment, this option must be selected at the beginning of the procedure. (See
instructions for “The Conventional Method”, Section 9.9).
• If the P
P
E
ER
RFFo
oR
R
M
MA
AX
XX
X®® G
Grriip
p is removed from the stent delivery system, it M
MU
US
ST
T N
No
oT
T be reattached. In this event, the stent M
MU
US
ST
T be
deployed using “The Conventional Method” of deployment.
• A removable red Safety Clip ((K
K)) prevents accidental or premature stent release.
• D
Do
o N
No
oT
T remove the Safety Clip ((K
K)) until you are ready to deploy the stent.
• Just prior to deploying the stent, the Safety Clip ((K
K)) must be removed by pressing the two red tabs ((LL)) together and removing the
clip from the grip. ((S
Seeee FFiig
gu
urree A
A2
2))
9
9..6
6
IIn
nttrrood
du
uccttiioon
n ooff tth
hee S
Stteen
ntt D
Deelliivveerryy S
Syysstteem
m::
• Insert the guidewire into the distal tip of the catheter until it exits the catheter at the proximal end of the device.
• Advance the delivery catheter over the guidewire into the target lumen. ((S
Seeee FFiig
gu
urree A
A5
5,, A
A6
6,, A
A7
7 aan
nd
d A
A8
8))
• Under fluoroscopic visualization, advance the stent delivery system across the stricture using the radiopaque markers to center the
stent across the lesion. ((S
Seeee FFiig
gu
urree A
A1
1))
• It is recommended to advance the delivery system past the stricture and then to pull back slightly on the entire system to achieve
the correct positioning of the markers and to help insure that slack has been removed and that the delivery catheter is straight.
• Prior to stent deployment, remove all slack from the catheter delivery system to avoid stent misplacement.
• D
Do
o N
No
oT
T hold the delivery system catheter during stent deployment. ((S
Seeee FFiig
gu
urree A
A3
3//A
A4
4))
9
9..7
7
S
Stteen
ntt P
Pllaacceem
meen
ntt::
• During stent deployment, the entire length of the catheter system should be kept as straight as possible. Maintaining a straight
catheter under slight tension during stent deployment is recommended to improve placement accuracy.
• Center the proximal stent markers (S
Seeee FFiig
gu
urree A
A1
1, “B2”) and both overlapping distal markers (S
Seeee FFiig
gu
urree A
A1
1, stent markers “B1”
and marker band on the outer catheter “C”) across the stricture. The radiopaque markers on the stent indicate the ends of the
compressed stent and the length of the expanded stent.
• By initially advancing the catheter beyond the stricture, micro-adjustments of the stent can be made by pulling the entire system
back toward the stricture to improve placement accuracy.
• once the stent is partially or fully deployed, micro-adjustments are no longer possible and the stent should N
No
oT
T be dragged or
repositioned in the lumen.
• once stent deployment has been initiated, the stent C
CA
AN
NN
No
oT
T be recaptured using the stent delivery system.
• once the moving marker has passed the proximal end of the stent by approximately 2 cm, the stent is completely deployed.
• Complete stent deployment can be fluoroscopically visualized when the radiopaque markers at the proximal and distal ends of the
stent are fully expanded.
9
9..8
8
S
Stteen
ntt D
Deep
pllooyym
meen
ntt W
Wiitth
h tth
hee P
P
E
ER
RFFo
oR
R
M
MA
AX
XX
X®® G
Grriip
p::
• There are three different stent deployment options with the P
P
E
ER
RFFo
oR
R
M
MA
AX
XX
X®® G
Grriip
p:
– “The Trigger Method” ((S
Seeee S
Seeccttiioon
n 3
3..3
3,, FFiig
gu
urree 1
1))
– “The Slide Method” ((S
Seeee S
Seeccttiioon
n 3
3..3
3,, FFiig
gu
urree 2
2))
– “The Combination Method (Trigger/Slide)”
• Switching from “The Trigger Method” to “The Slide Method” can be done at any time during stent deployment; however, switching
from “The Slide Method” to “The Trigger Method” M
MU
US
ST
T be avoided.
• D
Do
o N
No
oT
T remove the Safety Clip ((K
K)) until you are ready to deploy the stent.
• Just prior to stent deployment, remove the red Safety Clip ((K
K)) by pressing the two red tabs ((LL)) together and removing the clip from
the grip. ((S
Seeee FFiig
gu
urree A
A2
2))
• Under fluoroscopic visualization, deploy the stent utilizing your chosen method of deployment until the stent is fully deployed and
the slide mechanism has reached the proximal end of the handle. ((S
Seeee FFiig
gu
urree A
A9
9 –
– A
A1
11
1))
• During stent deployment ((S
Seeee FFiig
gu
urree A
A1
12
2)), the moving radiopaque marker (C) on the outer catheter moves backwards toward the
proximal markers on the stent (B2). The radiopaque markers on the stent (B1, B2) M
MU
US
ST
T N
No
oT
T move during stent deployment.
• After stent deployment, carefully withdraw the delivery system from the patient over the guidewire. After removing the delivery
system, visually confirm that the entire stent delivery system has been removed. ((S
Seeee FFiig
gu
urree A
A1
13
3))
(a) inner catheter
(b) outer catheter
(c) moving distal marker (C) on outer catheter
• Final radiological evaluation of the implanted stent should be conducted by angiogram.
9
9..9
9
S
Stteen
ntt D
Deep
pllooyym
meen
ntt U
Ussiin
ng
g T
Th
hee C
Coon
nvveen
nttiioon
naall M
Meetth
hood
d::
• In addition to the three deployment options outlined in Section 9.8 there is also an option to release the stent W
WIIT
TH
Ho
oU
UT
T the
P
P
E
ER
RFFo
oR
R
M
MA
AX
XX
X®® G
Grriip
p:
– “The Conventional Method” ((S
Seeee S
Seeccttiioon
n 3
3..3
3,, FFiig
gu
urree 3
3))
• To allow "The Conventional Method" of deployment, remove the white conversion tab ((M
M)) from the back of the grip. ((S
Seeee FFiig
gu
urree A
A1
14
4))
• Separate the P
P
E
ER
RFFo
oR
R
M
MA
AX
XX
X®® G
Grriip
p from the delivery system catheter by grasping the Luer lock ((E
E)) at the back end of the handle and
gently twisting to snap the catheter out of the back of the grip. ((S
Seeee FFiig
gu
urree A
A1
15
5)) Then grasp the deployment system at the front T-Luer
adapter ((FF)) and snap the deployment system completely out of the grip. ((S
Seeee FFiig
gu
urree A
A1
16
6)) Use caution not to bend the metal portion
of the catheter during removal from the grip.
• D
Do
o N
No
oT
T remove the Safety Clip ((K
K)) until you are ready to deploy the stent.
• Just prior to stent deployment, remove the red Safety Clip ((K
K)) by pressing the two red tabs ((LL)) together and removing the clip from
the grip. ((S
Seeee FFiig
gu
urree A
A1
17
7))
• Under fluoroscopic visualization, deploy the stent using the conventional “pin & pull-back” technique by slowly pulling back the T-Luer
adapter ((FF)) towards the hand that is pinned in place. Pulling back on the T-Luer adapter ((FF)) directly retracts the outer catheter and
deploys a corresponding portion of the stent. ((S
Seeee FFiig
gu
urree A
A1
18
8))
• Full stent deployment is ensured when the T-Luer adapter ((FF)) reaches the metal handle. ((S
Seeee FFiig
gu
urree A
A1
19
9))
• During stent deployment ((S
Seeee FFiig
gu
urree A
A1
12
2)), the moving radiopaque marker (C) on the outer catheter moves backwards toward the
proximal markers on the stent (B2). The radiopaque markers on the stent (B1, B2) M
MU
US
ST
T N
No
oT
T move during stent deployment.
• After stent deployment, carefully withdraw the delivery system from the patient over the guidewire. After removing the delivery
system, visually confirm that the entire stent delivery system has been removed. ((S
Seeee FFiig
gu
urree A
A1
13
3))
(a) inner catheter
(b) outer catheter
(c) moving distal marker (C) on outer catheter
• Final radiological evaluation of the implanted stent should be conducted by angiogram.
9
9..1
10
0 P
Poosstt--S
Stteen
ntt P
Pllaacceem
meen
ntt::
• Post-dilatation of the stent with an appropriately sized balloon dilatation catheter is left to the discretion of the treating physician.
• The B
B
A
AR
RD
D
®® E
E
LL
U
UM
MIIN
NE
EX
XX
X
®® V
Vaassccu
ullaarr S
Stteen
ntt is a self-expanding, nitinol stent that M
MU
US
ST
T N
No
oT
T be expanded beyond its labeled diameter by
dilatation with a PTA balloon.
• This product has been designed for single patient use only. D
Do
o N
No
oT
T reuse. D
Do
o N
No
oT
T resterilize.
• After use, the stent delivery system is a potential biohazard. Handle and dispose of this product in accordance with accepted medical
practice and with applicable local, state and federal laws and regulations.
1
10
0..0
0 P
PA
AT
TIIE
EN
NT
T IIM
MP
PLLA
AN
NT
T IIN
NFFo
oR
RM
MA
AT
TIIo
oN
N C
CA
AR
RD
DS
S
• A Patient IMPLANT Information Card is provided in the back of the IFU for your convenience.
8
BAWB05657:BAWB05657
08.03.2010
15:51
Seite 8
Содержание E-Luminexx Vascular Stent
Страница 79: ...79 BAWB05657 BAWB05657 08 03 2010 15 52 Seite 79...
Страница 80: ...80 BAWB05657 BAWB05657 08 03 2010 15 52 Seite 80...
Страница 94: ...BAWB05657 BAWB05657 08 03 2010 15 52 Seite 94...
Страница 96: ...BAWB05657 BAWB05657 08 03 2010 15 52 Seite 96...
Страница 98: ...BAWB05657 BAWB05657 08 03 2010 15 52 Seite 98...
Страница 100: ...BAWB05657 BAWB05657 08 03 2010 15 52 Seite 100...
Страница 102: ...BAWB05657 BAWB05657 08 03 2010 15 52 Seite 102...