38
The ProVOLT function allows for maintaining, stepping up or stepping down the
voltage output between steps without the voltage dropping to 0V. If ProVOLT = ON,
the output voltage will not drop to 0V between test steps.
Additional notes on the ProVOLT function:
- Test steps must be of the same test type (ACW, DCW, IR or
–DCW). For
example if an ACW test is followed by a DCW test, the ProVOLT function will
not operate.
- If stepping up the voltage between steps, the ramp down parameter is ignored.
- If stepping down the voltage between steps, the ramp up parameter is ignored.
You can set The ProVOLT function to ON or OFF. To set the ProVOLT function touch
the ON/OFF key. To save the setting touch the enter key (
↵
)
and move to the next
parameter.
4.4.7 Barcode
From the Setup System menu, touch the Barcode icon and the following screen will
appear: