141
COMMAND
DESCRIPTION
VALUE
SP
Step Prompt Delete
LP?
List Prompt
LP nn
List Prompt current
(1) "Valid ASCII" is the character set that is available from the front panel LCD user
interface. Consisting of upper case alphabet (A-Z), numbers (0-9) and decimal point
(.), asterisk (*), dash (-), under bar (_), tilde (~) and space (SP).
FL <memory number>
Load a file by memory number from non-volatile memory into random access memory
RAM.
FN < file name>
Creates a new file name for the active memory loaded into RAM.
SS <step number>
Selects the active selected step to load into RAM. The step must first be selected
before any specific parameters can be edited.
SAA, SAD, SAI, SAC
These commands add the appropriate test type within the memory at the step location
that has been selected. The parameters of the previous test type will be deleted and
the default values for the new test type will be recalled. If the same test type is
selected that already exists, the default values will replace the previous parameters.
ADD2 <n, p1,p2,p3…>
This command adds a step with all parameters to the current file. See the command
summary tables below to see the specific test type for each of these commands
The parameter <
n>
indicates the test type. The values ACW, DCW, IR, or CONT.
must be used. The parameters <
p1,p2>
etc. indicate the individual settings for each
parameter of the test. All parameters must be included with the command and should
appear in the same order that is shown in the table below. Also, like the individual
parameter editing commands, the unit should not be included with the value; only the
numeric value should be included in the command string. When the scanners are
being used they should be appended to the end of the string, with the internal scanner
first if installed, followed by the external scanner if connected to the rear panel
scanner control port.
The list of parameters can also be found in the default parameters section of the
manual, or refer to
Section 7.4.5
Test Parameter Editing Commands and
Companion Queries
for the proper values.
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...