53
Select the value and the enter key (
↵
)
will appear. Touch the enter key to save the
value and move to the next parameter under the Calibration Alert menu. Use
the Prev
and Next arrows to toggle between different fields under the Calibration Alert menu.
Calibration Due Date and Alert Date
We recommended that calibration is performed at least once a year. We also
recommend that the Calibration Due Date is not set greater than one year from the
Calibration Date displayed. After a calibration is performed, the instrument
automatically sets the Calibration Due Date 1 year after the calibration date.
This parameter defaults to one year after the calibration date but you may overwrite to
any date desired.
You can set the Calibration Due Date and Alert Date by selecting the appropriate
fields on the touchscreen. The following screens will appear when you select the
Calibration Due Date and Alert Date fields:
The Alert Date is like an alarm clock that will warn you in advance of the actual
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...