79
4.12.2 Default Test Parameters
Each of the test types (AC Withstand, DC Withstand, Insulation Resistance and
Continuity) have specific default test parameters that automatically load when you
select the test from the Parameter Edit screen. The following table is a listing of the
Default Parameters for each of the different types of tests that are available in the
HypotULTRA. The Default
Parameters are as follows:
TEST TYPE PARAMETER
VALUE
ACW
Voltage
1240V
HI-Limit T
10.00mA
LO-Limit T
0.000mA
HI-Limit R
10.00mA
LO-Limit R
0.000mA
Display
Total
Ramp UP
0.1s
Dwell
1.0s
Ramp DN
0.0s
Frequency
60Hz
Arc Detect
OFF
Arc Sense
5
Continuity
OFF
Scanner
OOOOOOOO
Prompt
DCW
Voltage
1500V
(7850 and
7800 only)
HI-Limit
10000uA
LO-Limit
0.0uA
Ramp UP
0.1s
Dwell Time
1.0s
Ramp DN
0.0s
Ramp-HI
OFF
Charge-LO
0.0uA
Arc Detect
OFF
Arc Sense
5
Continuity
OFF
Scanner
OOOOOOOO
Prompt
IR
Voltage
500V
(7850 and
7800 only)
HI-Limit
0.00M
LO-Limit
0.05M
Ramp UP
0.1s
Delay
1.0s
Ramp DN
0.0s
Charge-LO
0.000uA
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...