4.3 Setup System
From the default main menu touch the Setup System icon to enter the Setup System
sub menu. This is where all System Parameters are edited:
Default System Parameters
HypotULTRA comes from the factory with the following system presets:
SYSTEM PARAMETERS
Setup
Sys.
PLC Remote
OFF
Single Step
OFF
Fail Stop
ON
Alarm
5
Results
Last
Address (GPIB only)
8
Smart GFI
ON
Results
Last
Cal Alert
ON
Security
Password from Factory
0
Security based on Password
OFF
Memory, Step, and Default System Parameter Initialization
Initializing the instrument will overwrite all memories and
steps with ACW default parameters!
Initialisation de l’instrument va écraser toutes les mémoires et les étapes avec les
paramètres par défaut ACW
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...