36
appear:
You can set the Single Step function to ON or OFF. To set the Single Step, touch the
ON/OFF key. To save the setting touch the enter key (
↵
)
and move to the next
parameter.
4.4.4 Fail Stop
The next parameter in the Setup System menu is Fail Stop. Fail Stop is a function that
will stop a sequence if a failure occurs. If this function is turned off, the sequence of
tests will continue to the end of the sequence regardless of whether or not a failure
has occurred. If the Fail Stop is OFF and a failure occurs during the test sequence,
the RESET button will light and a short alarm will sound but the sequence will
continue to the end.
At the end of the test sequence the RESET button will light and alarm will sound
indicating failure during the sequence. Pressing the RESET button will silence the
alarm. Pressing the RESET button a second time will reset the instrument. From the
main Setup System menu touch the Fail Stop icon and the following screen will
appear: