178
COMMAND
NAME
TEST
TYPES
VALUE
EHT <
value
>
EHT?
Edit Max-Lmt-T
ACW
0.000 - 30.00mA
EHR <
value
>
EHR?
Edit Max-Lmt-R
ACW
0.000 - 30.00mA
EH <
value
>
EH?
Edit Max-Lmt
DCW
IR
CONT
0.0 - 10000uA
0, 0.05 - 50000M
0, 0.00 - 2000
ELT <
value
>
ELT?
Edit Min-Lmt-T
ACW
0.000 - 30.00mA
ELR <
value
>
ELR?
Edit Min-Limit-R
ACW
0.000 - 30.00mA
EL <
value
>
EL?
Edit Min-Lmt
DCW
IR
CONT
0.0 - 10000uA
0.05 - 50000M
0.0 - 2000
EF {1|0}
EF?
Edit Frequency
ACW
1=60Hz, 0=50Hz
EIS
<scanner
string>
EIS?
Edit Internal Scanner
ACW
DCW
IR
CONT
scanner string
= 1-16 element
ASCII string consisting of H, L, or
O.
H=HV or Cont. , L=RETURN,
O=OPEN
EES
EES?
Edit External Scanner
ACW
DCW
IR
CONT
scanner string
= 1-16 element
ASCII string consisting of H, L, or
O.
H=HV or Cont., L=RETURN,
O=OPEN
System Parameter Editing Commands and Companion Queries
These commands are used to modify the system parameters for the instrument.
These commands require a parameter value to be included with the command. The
companion query command will read the parameter using the same value that is used
for setting the parameter.
COMMNAD
NAME
VALUE
SPR {1|0}
SPR?
PLC Remote
1= On, 0=Off
SSI {1|0}
SSI?
Single Step
1= On, 0=Off
SF {1|0}
SF?
Fail Stop
1= On, 0=Off
SAL <
value
>
SAL?
Alarm Volume
0-9
SC <
value
>
SC?
Contrast
1-9
SL {1|0}
Lock
1= On, 0=Off
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...