124
6.6
Error Messages
Abort:
This message appears on the display if you abort the test in process with the
RESET button or remote Reset control.
HI-Limit:
This message appears on the display if the DUT measurement exceeds the
HI-Limit setting of any parameter.
LO-Limit:
This message appears on the display if the DUT measurement drops below
the LO-Limit
Max-Lmt:
This message appears on the display if the DUT measurement exceeds the
Max-Lmt setting of any parameter.
Min-Lmt:
This message appears on the display if the DUT measurement drops below
the Min-Lmt.
CONT-Fail:
This message appears on the display if the DUT fails the basic continuity
check performed during an AC/DC Withstand
test (if Continuity is selected “ON”).
Arc-Fail:
This message appears on the display if the DUT arcing current exceeds the
Arc
Sense limit and Arc function is active (Arc Sense = 1…9) of the AC/DC Withstand
test.
Short:
This message appears on the display if the DUT current is well beyond the
metering range of the test.
Charge-LO:
This message appears on the display if the leakage current during
Ramp-up falls below the Charge -LO setting.
Breakdown:
This message appears on the display if the DUT current is well beyond
the metering range of the test and the arcing condition beyond the arc sense limit.
GND-Fault:
This message appears on the display if the GFI threshold is exceeded
during the test.
Interlock Open:
This message appears on the display if the Remote Interlock feature
is activated before or during a test. The Remote Interlock feature utilizes a set of
closed contacts which will disable the instrument’s output if they are opened before or
during a test. You can also refer to the Remote Interlock as a remote system lockout,
utilizing “fail when open” logic. You can disable the Remote Interlock feature by
plugging the “Interlock Disable Key” provided into the Signal Input connector.
Output Error:
This message appears on the display, if the instruments output reading
does not match the setting. You will only see this message if you press the EXIT key
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...