152
COMMAND
NAME
DESCRIPTION
*CLS
Clear Status Command
Clear Standard Event Status
Register
Clear Service Request Register
*OPC
Operation Complete Command
When all selected pending
operations complete, ESR BIT0=1
*OPC?
Operation Complete Query
When all selected pending
operations complete, Output
Queue=1
*WAI
Wait-to-Continue Command
*PSC {1|0}
Power-on Status Clear Command 1 = Power-on clear enable
registers
0 = Power-on load previous enable
registers
*PSC?
Power-on Status Clear Query
*ESR?
Standard Event Status Register
Query
0 - 255
*ESE <
value
> Standard Event Status Enable
Command
value
= 0 - 255
*ESE?
Standard Event Status Enable
Query
0 - 255
*STB?
Read Status Byte Query
Read Status Byte
*SRE
<
value
>
Service Request Enable Command
value
= 0 - 255
*SRE?
Service Request Enable Query
0 - 255
*IDN?
Read the instrument identification string. Company = Associated Research Inc.
*RST
Reset the instrument to original power on configuration. Does not clear Enable
register for Standard Summary Status or Standard Event Registers. Does not clear
the output queue. Does not clear the power-on-status-clear flag.
*TST?
Performs a self-test of the instrument data memory. Returns 0 if it is successful or 1 if
the test fails.
*CLS
Clears the Status Byte Summary register and Event registers. Does not clear the
Enable registers.
*OPC
Sets the operation complete bit (bit 0) in the Standard Event register after a command
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...