176
0.00 - 10.00
0.01
± (2% of r 2 counts)
Additional error when reading <
6% of range.
± (0.2% of range)
13 400/800 Hz Hipot Output
Perform ACW test at 400Hz or 800Hz.
14 Mimic HU3 Command Set
The Mimic HU3 Command Set option for the new HypotULTRA allows you to emulate
the HypotULTRA 3 making it a drop-in replacement for the older model. The new
interface commands and structure of the HypotULTRA can emulate those of the
HypotULTRA 3 interface. This will allow HypotULTRA 3 users to integrate the new
model into existing test systems easier. The following command tables and
descriptions provide the commands that can be used with this option:
Test Execution Commands
The following commands are used to control actual output voltage and current from
the instrument. Please observe all safety precautions.
COMMAND
DESCRIPTION
TEST
Execute a Test
RESET
Abort a test in Process or Reset Failures
SAO
Set Auto-Offset
SACG
Set Auto-Charge-LO
File Editing Commands
The following commands are used to create or modify Test Setup Files.
Commands should be separated from parameters by a space.
If multiple parameters are entered, they should be separated by commas.
COMMAND
DESCRIPTION
VALUE
FL <
memory number
>
File Load
memory number
= 1-50
FN <
file name
>
File Name
file name =
Valid ASCII (1)
maximum 11 characters
SS <
step number
>
Step Select step number
step number
= 1-30
SAA
Step Add ACW test
SAD
Step Add DCW test
SAI
Step Add IR test
SAC
Step Add CONT test
ADD
<test,
p1,p2,p3…>
Add all parameters of a
test
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...