94
There are two options for setting the frequency of the applied voltage for an ACW test.
Use the touchscreen keypad to select between 50Hz or 60Hz. Use the enter key (
↵
)
to
save the entry and move to the next test parameter, Continuity:
The Continuity function within an ACW test is a basic ground continuity check that can
be set to ON or OFF. Use the touchscreen keypad set the Continuity function and use
the enter key (
↵
)
to save the entry and move to the next test parameter, Range: