144
DCW Test
T
es
t
V
o
lt
age
Hi
-L
imit
LO
-L
imit
Ramp
Up
Dw
ell
T
ime
Ramp
Do
w
n
Ch
ar
g
e LO
Arc
S
ens
e
O
ff
se
t
Ramp
-Hi
Arc
D
etect
Co
n
tinu
it
y
Rang
e
L
o
w
Rang
e
DCW
1500
10000
0.0
0.4 1.0
0.0
0.0 5
0.0
0.0
OFF
OFF AUTO
OFF
ADD2 DCW,1500,10000,0.0,0.4,1.0,0.0,0.0,5,0.0,0.0,OFF,OFF,Auto,OFF
Insulation Resistance Test
T
es
t
V
o
lt
age
Hi
-L
imit
Lo
-L
imit
Ramp
Up
Dela
y
T
ime
Dw
ell
T
ime
Ramp
Do
w
n
Ch
ar
g
e
-
LO
IR
500 0.00 0.10 0.1 0.5
0.5
0.0
0.000
ADD2 IR,500,0.00,0.10,0.1,0.5,0.5,0.0,0.000
Continuity Test
T
es
t
Max
-Lmt
Min
-Lmt
Dw
ell
T
ime
O
ff
se
t
CONT. 1000 0.000 1.0
0.00
ADD2 CONT.,1000,0.000,1.0,0.00
Use the following example command string with tester that includes optional scanners
channels. The BUS command follows similar structure to the ADD2 command string
mentioned above.
ADD2 ACW,1240,10.00,0.000,0.1,1.0,0.0,5,10.00,0.000,0.000,60,OFF,OFF,Auto,OOOOOOOOOOOOOOOO
ADD2 DCW,1500,10000,0.0,0.4,1.0,0.0,0.0,5,0.0,0.0,OFF,OFF,Auto,OFF,OOOOOOOOOOOOOOOO
ADD2 -DCW,-1500,-10000,0.0,0.4,1.0,0.0,0.0,5,0.0,0.0,OFF,OFF,Auto,OFF,OOOOOOOOOOOOOOOO
ADD2 CONT.,1000,0.000,1.0,0.00,OOOOOOOOOOOOOOOO
ADD2 IR,500,0.00,0.10,0.1,0.5,0.5,0.0,0.000,OOOOOOOOOOOOOOOO
SP <prompt message>
Adds or edits a prompt message for the active step.
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...