155
After the serial poll (SPOLL) is executed the RQS bit will be cleared to 0, and the
remaining bits will remain unchanged. The Status Byte will not change value until the
event register is read and cleared for the corresponding Status Byte bit.
For example after the All Pass SRQ has been enabled, when the test(s) have finished
with pass indications the instrument will set the hardware SRQ line and output the
Status Byte of 41 hex. This means that bit 6 and bit 0 are set to a value of 1. After
reading the Status Byte the Status Byte value will change to 01 hex.
FOR MORE INFORMATION ON IEEE (GPIB) PLEASE CONTACT
The Institute of Electrical and Electronic Engineers, Inc.
345 East 47th Street,
New York, NY 10017
1-212-705-7018 (Communications Society of IEEE)
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...