150
COMMAND
NAME
VALUE
LF?
List File Name
Active selected
memory
LF nn?
List File Name
Active selected file
LP?
List Prompt
Active selected Step
LP <
step
number
>?
List Prompt by step number
step number
= 1-200
LS2?
List Step Parameters
LS2 <
step
number
>?
List Step Parameters by step
number
step number
= 1-200
TD?
Read the active data being displayed on the LCD display while the test is in process.
Also read the last data taken when the test sequence has completed. Each
parameter is separated by commas and includes step number, test type, test status,
and metering. The syntax for this command response is {step, test type, status, meter
1, meter 2, meter 3}. ACW test displays 4 meters. Each meter will contain only the
value and not the units. In the case of DCW current where both uA and mA are used
on the display, the command response will always indicate the current in uA for
example 2.0mA will respond with 2000 for 2000uA.
RD <step number>?
Read the results for an individual step. The step number is the actual step number
that has been saved within the file, not the order of which the steps were executed.
For example if the test was executed starting from step 3 and ending with step 5 then
the first step test results will be found in location 3 not in location 1. Each parameter
is separated by commas and includes step number, test type, test status, and
metering. The syntax for this command response is {step, test type, status, meter
1,meter 2,meter 3}. ACW test displays 4 meters. Each meter will contain only the
value and not the units. In the case of DCW current where both uA and mA are used
on the display the command response will always indicate the current in uA for
example 2.0mA will respond with 2000 for 2000uA.
RR?
Read the remote Reset input signal. When the remote reset has been activated by
closing the contacts the query will return a value of 1 to indicate the instrument is
being Reset.
RI?
Read the remote Interlock input signal. When the remote Interlock has been activated
by opening the contacts the query will return a value of 0 to indicate the instrument is
in the Interlock state and will not be able to generate output voltage or current.
RS?
Read Scanner Status command will respond with a value that identifies the number of
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...