177
COMMAND
DESCRIPTION
VALUE
SP <
prompt message
>
Step Prompt Create
prompt message
= Valid ASCII (1)
maximum 32 characters
SP
Step Prompt Delete
SF {1|0}
Step Fail Stop
1=On, 0=Off
Test Parameter Editing Commands and Companion Queries
These commands are used to modify the test parameter within each step. These
commands require a parameter value to be included with the command. The
companion query command will read the parameter. The writing of the parameter
requires that the unit not be included with the value, only the numeric value should be
included with the command. Also, when the query commands are used the response
will not include the unit’s characters. Many of the commands will function the same
way for multiple test types; however, the input range may be different and therefore
used a different possible set of values.
COMMAND
NAME
TEST
TYPES
VALUE
ECC {1|0}
ECC?
Edit Connect
ALL
1= On, 0=Off
ED {1|0}
ED?
Edit Current Meter
Display
ACW
1 = Real, 0 = Total
EV <
value
>
EV?
Edit Voltage
ACW
DCW
IR
0 - 5000V
0 - 5000V
50 - 1000V
ECG <
value
>
ECG?
Edit Charge-Lo
DCW
IR
0.0 - 350.0uA
0.000 - 3.500uA
ECT {1|0}
ECT?
Edit Continuity
ACW
DCW
1= On, 0=Off
ERU <
value
>
ERU?
Edit Ramp-Up
ACW
DCW
IR
0.1 - 999.9s
ERD <
value
>
ERD?
Edit Ramp-Down
ACW
DCW
IR
0.0 - 999.9s
0, 1.0 - 999.9s
0, 1.0 - 999.9s
ERH {1|0}
ERH?
Edit Ramp-HI
DCW
1= On, 0=Off
EDW <
value
>
EDW?
Edit Dwell
ACW
DCW
CONT
0, 0.4 - 999.9s
0, 0.4 - 999.9s
0, 0.3 - 999.9s
EDE <
value
>
EDE?
Edit Delay
IR
0, 1.0 - 999.9s
EO <
value
>
EO?
Edit Offset
CONT
0.00 - 10.00
EA <
value
>
EA?
Edit Arc
ACW
DCW
1 - 9
EAD {1|0}
EAD?
Edit Arc-Detect
ACW
DCW
1= On, 0=Off
Содержание HypotULTRA 7800
Страница 25: ...20 0 00 10 00 0 01 2 of reading 2 counts Additional error when reading 6 of range 0 2 of range ...
Страница 34: ...29 ...
Страница 53: ...48 ...
Страница 121: ...116 ...
Страница 128: ...123 ...
Страница 143: ...138 ...
Страница 187: ...182 ...
Страница 196: ...191 Calibration of 9 99mA DC Current Range The 10mA DC current range calibration is a one step process ...
Страница 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...