4.4 Hardware
The Hardware menu contains important system parameters which you must set prior
to performing tests. Touch the Hardware icon and the following screen will appear:
You can scroll the touchscreen downward to see the remaining Hardware parameters.
Upon scrolling down the remaining items appear as follows:
4.4.1 Smart GFI
The high voltage power supply of the HypotULTRA is internally referenced to earth
ground. Since the leakage current measuring circuit of the instrument monitors only
current that flows through the return lead the possibility exists for current to flow