OPERATION
34
5. If the resistance exceeds the high limit trip setting and exceeds the metering range, the
red “FAIL” symbol will illuminate and the alarm will sound. The display will show:
HI - LMT MX XXX . X s
XX . XX A OFL m
Ω
6. If the resistance does not exceed the low limit setting, the red “FAIL” symbol will
illuminate and the alarm will sound. The display will show:
LO - LMT M X XXX . X s
X X . XX A XXX m
Ω
7. To stop the alarm, please press the “RESET” switch once. The alarm will stop and
the display will retain the failure information. If the “Reset” switch is pressed again, the
data on the display screen will be cleared and the display will indicate the setting data from
the last test. The instrument is now ready for the next test.
8. If the DUT passed all of the tests, the green “PASS”
”
symbol on the front panel will
illuminate and the instrument will output a short audible beep tone to indicate the DUT
has passed the tests. The instrument is now ready to perform another test.
The display will show:
Pass MX XXX . X s
XX . XX A XXX m
Ω
9. If the operator elects to abort a test in process this can be accomplished by pressing
the “Reset” switch at anytime. The instrument will stop the test process immediately.
The display will show:
Abort MX XXX . X s
XX . XX A XXX m
Ω
Please press the “Test” switch to initiate another test.