F. ADAPTER BOX CONNECTION
The adapter box is an optional accessory and is not provided as standard equpment with
the HYAMP II. If you would like to aquire a 36541 adapter box for use with your
HYAMP II please contact Associated Research using the contact information provided in
the Safety section of this manual.
The following diagram shows how to connect the optional adapter box to the HYAMP II
and the Device Under Test. This adapter box allows the user an easy way to connect an
item which is terminated in a three prong line cord. The high current is wired to the
ground pin of the receptacle box and from there the test is performed on the ground
conductor of the DUT to the chassis or dead metal of the product. The high voltage lead
is not used and should be secured out of the way.
If the HYAMP II and the HypotPLUS II are being used as a test system, the high voltage
lead should be plugged into the high voltage jack on the front panel of the HypotPLUS II.
The return lead from the HypotPLUS II should be connected to the HYAMP II as
described in the previous section (Common Cable Connections). This configuration would
allow you to do both the ground bond and hipot tests by pushing a single test button on
the HYAMP II, once all connections are made and the test parameters are set up.
(Optional)
5030DT
H.V.
Test Connections
36541
DUT
CHASSIS
L
N
E
CURRENT
RETURN
Содержание HYAMP II 5030DT