OPTIONS
43
fail condition will remain until the resistance drops below the Hi-Limit setting, or until the
Reset button is pressed to deenergize the output.
This operation emulates the basic audible continuity testers found on many multimeters,
and is useful in doing multipoint ground bond tests on the DUT by disconnecting and
reconnecting the test leads without disabling the output. This method is perfectly safe to
the operator with regards to the voltage potential that is available on the test leads. This
open circuit voltage is controlled by the voltage setting parameter. The user can access
the voltage setting parameter by pressing the VOLTAGE button on the left hand side of
the display on the front panel. The maximum voltage setting available is 8 volts AC and
presents absolutely no risk to the operator.