QUICK START
23
a). Unpack this instrument from its special shipping container. Be sure to save all
packaging materials in case you need to return it to the factory for service.
b). Locate a suitable testing area and be sure you have read all safety
instructions for the operation of the instrument and suggestions on the
test area set-up in the SAFETY section of this manual. Locate a three prong grounded
outlet. Be sure the outlet has been tested for proper wiring before connecting the
instrument to it.
c). Check to be sure the correct input line voltage has been selected on
the rear panel. Either 115 volts AC or 230 volts AC. Connect the power
input plug into its socket on the rear panel of the instrument. Connect the male end of the
plug to the outlet receptacle.
Please be sure that the safety ground on the power line cord is not
defeated and that you are connecting to a grounded power source.
d). Turn on the POWER switch located on the lower left hand side of the front panel.
Upon powering the instrument up a POWER ON SELF TEST (POST) will automatically
be performed. This test will check for the condition of the ram chips, led indicators, pcb's
and other critical components. All of the front panel led's will light temporarily then go
out. In addition, you will see the Associated Research name and Model Number briefly
appear on the LCD readout and then clear itself.
ARI HYAMP II
5030 DT VER
You should then see the default parameters on the LCD meter as follows:
Set M1 1.0s
25.00A 100m
Ω
These abbreviated parameters stand for the following:
SET:
This is the parameter settings review screen.
M1:
The instrument is using the default set ups contained in memory 1.
1.0s:
The dwell timer is set to a test duration of 1 second.
25.00A:
The test current is set to 25.00 Amps.
100m
Ω
:
The high resistance trip point is set to 100 milliohms.
WARNING
CAUTION
Current software
version displayed here.