QUICK START
25
QUICK GUIDE TO INTERCONNECTING HYAMP II TO HypotPLUS II
USING THE SUPPLIED CABLE KIT
1. Enable the HypotPLUS II remote interface. Operation menu REMOTE = "ON".
2. Connect the Double ended banana plug cable from the HypotPLUS II rear RETURN
jack to the HYAMP II INTERCONNECT jack.
3. Based on the test method, select one of the two (9 pin dsub) control cables shown in
the table below. Connect this cable along with the Common control cable (9 pin dsub
5030DT-21) between the HYAMP and the HypotPLUS II remote I/0 with the
BLUE
ends on the HYAMP II
.
Note: The cables are identified on the blue tubing.
Cable number
Method
Operation
5030DT-22
Sequential
HYAMP first then Hypot
5030DT-23
Simultaneous
HYAMP and Hypot at the same time
4. Connect the high current plug (red) from the receptacle adapter box (P/N 36541) to
the HYAMP II jack marked CURRENT (red), and the high voltage plug (white) to the
HypotPLUS II jack marked HV (white).
5. Connect the RETURN lead with the alligator clip (black)
to the HYAMP II jack
marked RETURN (black). The alligator clip should be connected to the DUT chassis.
6. Check the test parameters and initiate the tests by pressing the Hyamp test button. The
Reset buttons on both instruments are active to disable the tests or for failure reset.
•
For a more detailed description on interconnecting your system, see page 37.