SPECIFICATIONS
8
Model 5030DT Functional Specifications
GROUND BOND TEST MODE
Unless otherwise stated, accuracies are relative to a laboratory standard measurement.
Descriptions Specifications
INPUT VOLTAGE
115 / 230V selectable,
±
15
%
variation
47 - 63 Hz
FUSE
115 VAC, 230VAC -- 5A fast acting 250VAC
OUTPUT
Current:
AC 3 - 31 Amps, 0.01 Amp / step
Regulation: +(2% of S 0.02 A)
Voltage: AC 3 - 8 Volts, 0.1 Volt / step
OUTPUT
FREQUENCY
50 / 60 Hz selectable
DWELL TIME
SETTING
0 and 0.5 - 999.9 seconds, 0.1 second / step
“0” for continuous running
Accuracy: +(0.1% of S 0.05 seconds)
FAILURE SETTINGS High limit: 0 - 600m
Ω
for 3 - 12 Amps, 1m
Ω
/
step
0 - 150m
Ω
for 3 - 31 Amps, 1m
Ω
/ step
Accuracy: +(2
%
of setting
+
2m
Ω
)
Low limit: 0 - 600m
Ω
for 3 - 12 Amps, 1m
Ω
/
step
0 - 150m
Ω
for 3 - 31 Amps, 1m
Ω
/
step
Accuracy:
±
(2
%
of setting
+
2m
Ω
)
MILLIOHM
OFFSET
Max. Offset Capability: 100m
Ω
, 1
m
Ω
/ step
Accuracy:
±
(2
%
of setting
+
2m
Ω
)
METERING
Ammeter (4 digits)
Range:
0 - 31 Amps
Resolution: 0.01 Amp / step
Accuracy: +(3% of R 0.03A)
Ohmmeter (3 digits)
Range:
0 - 600m
Ω
Resolution: 1 m
Ω
/ step
Accuracy: +(2% of R2m
Ω
)
Timer (4 digits)
Range:
0 - 999.9 seconds
Resolution: 0.1 seconds / step
Accuracy: +(0.1% of R 0.05 seconds)