OPERATION
28
2. To set the output test voltage
Please press the “VOLTAGE” key, the display will show:
V o l t a g e = X.XX V
Use the “Numerical” keys to enter the voltage setting, then press the “Enter” key to store
the setting. The unit of measure is in volts with 3.00 - 8.00 being the range.
3. To set the high limit resistance trip setting
Please press the “HI-LIMIT” key, the display will show:
HI-LMT = XXX m
Ω
Please use the “Numerical” keys to enter the high limit resistance trip setting, then press
the “Enter” key to store the setting. The unit of measure is in milliohms with a range from
0-600 m
Ω
’s.
4. To set the low limit resistance trip setting
Please press the “LO-LIMIT” key, the display will show:
LO-LMT = XXX m
Ω
Please use the “Numerical” keys to enter the low limit resistance trip setting, then press
the “Enter” key to store the setting. The unit of measure is in milliohms with a range from
0-600 m
Ω
’s, if set for “0”, the low limit resistance trip setting will be disabled.
5.
To set the dwell time
Please press the “DWELL” key the display will show:
Dwell - T = XXX . X s
Use the “Numerical” keys to enter the dwell time setting, then press the “Enter” key to
store the dwell time setting. The range is 0 to 999.9 seconds, if the dwell time is set to
“0”, the instrument will operate in a continuous ON mode when the test button is