![Associated Research HYAMP II 5030DT Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/associated-research/hyamp-ii-5030dt/hyamp-ii-5030dt_operation-and-service-manual_2989497050.webp)
CALIBRATION
46
CALIBRATION PROCEDURES
This instrument has been fully calibrated at the factory in accordance to our published
specifications. It has been calibrated with standards traceable to NIST. You will find in
this manual a copy of the "Certificate of Calibration". It is recommended that you have
this instrument recalibrated and a safety check done at least once per year. AR
recommends you use "Calibration Standards" that are NIST traceable, or traceable to
agencies recognized by NIST to keep this instrument within published specifications. End
user metrology standards or practices may vary. These metrology standards determine the
measurement uncertainty ratio of the calibration standards being used. Calibration
adjustments can only be made in the Calibration mode, calibration checks can only be
made in the Test mode of operation.
Calibration Equipment Required:
The following equipment will be needed to properly calibrate your instrument.
1). Digital Multimeter with the minimum ranges;
ac voltage: 10 VAC
ac current: 30 AAC.
CALIBRATION PROCEDURE:
To enter the calibration mode the instrument must be in the OFF position. Press and hold
the “0” key, then turn on the Input Power Switch. The display will show
:
C a l i b r a t i o n M o d e
0 : V 1: A 7 : L o c k
Press “0” for volts calibration, “1” for amps and “7” for the password and lock mode
settings.
1. Calibration of Voltage
Equipment needed: STANDARD AC VOLTMETER capable of measuring 10VAC.
Please connect a standard ac voltmeter with 10V minimum full scale range to the current
and return connectors. Then press the “0” key on the front panel. The instrument will
provide approximately 7.5VAC on the output connectors and the display will show:
V o l t a g e = __ V