CALIBRATION
47
Please use the numerical keys to enter the reading from the ac voltmeter into the
instrument. Then press the “Enter” key to store the voltage setting. If an incorrect value is
entered, press the “EXIT” key. Then the correct value may be re-entered.
2. Calibration of Current
Equipment needed: STANDARD AC AMMETER with a range of 30A.
Please connect the standard ac ammeter across the output leads of the instrument. Press
the “1” key. The instrument will provide around 30A on the output and the display will
show:
C u r r e n t = __ A
Please use the numerical keys to enter the reading of the standard ac ammeter into the
instrument. Then press the “Enter” key to store the current setting. If an incorrect value
is entered, press the “EXIT” key. The correct value may be re-entered.
3. Setup of the Password and Lock Function
Please refer to the setup procedures section of this manual.
4. Exit Calibration Mode
After the calibration is complete, turn the input power switch to OFF (0) to exit the
Calibration Mode. Turn the power switch to ON (|) to return to standard operating test
mode for calibration verification.