CONTROLS
18
1. INTERCONNECT JACK:
For the connection of the return lead from the hipot
tester used when performing both hipot and Ground Bond tests on the same test item.
2. RETURN OUTPUT JACK:
For the connection of the detachable 5 foot (1.52 m)
black kelvin return test lead. This lead is always used when performing a test.
3. CURRENT OUTPUT JACK:
For the connection of the detachable 5 foot (1.52 m)
red kelvin test lead. This lead is always used when performing a ground bond test.
4. THERMAL
FAN:
Cycles on to cool the instrument automatically.
5. REMOTE
OUTPUT:
9 pin “D” subminiature female connector for monitoring
PASS, FAIL, and PROCESSING output relay signals, as well as remote START and
RESET outputs.
6. REMOTE
INPUT:
9 pin “D” subminiature male connector for remote control of test
and reset funtions as well as program memory selection 1, 2, or 3 and withstand
processing input.
7. CHASSIS GROUND (EARTH) TERMINAL:
This safety terminal should be
connected to a good earth ground before operation.
8. INPUT POWER SWITCH:
Line voltage selection is set by the position of the
switch. In the down position it is set for 115 volt operation, in the up position it is set
for 230 volt operation.
9. INPUT POWER RECEPTACLE:
Standard IEC 320 connector for connection to
a standard NEMA style line power (mains) cord.
10. FUSE RECEPTACLE:
To change the fuse unplug the power (mains) cord and turn
the fuse cap counter clockwise to remove the fuse.
REAR PANEL CONTROLS