QUICK START
24
If you wish to not use any one of these parameters you must overwrite the memory 1
position or change your parameters and save them in a different memory such as memory
position 2.
e). If the instrument defaults are acceptable then be sure to connect the appropriate test
leads to the device under test (DUT) or test fixture. Be sure to connect this safety ground
to a suitable known good ground before energizing this instrument, Then connect the
return lead first (black) to the test fixture or item followed by the high current output lead
(red).
f). Please check your connections to be sure they are making good
contact and that the test station or area is clear of debris and other
personnel. DO NOT TOUCH THE DEVICE UNDER TEST ONCE THE TEST HAS
BEEN STARTED. To initiate the test press the GREEN test button on the front panel.
This is a momentary button and does not need to be held in the pressed position during the
test. The instrument will then cycle ON and begin the automated test sequence using the
defaults. If a failure occurs you will HEAR an audible alarm go off. To stop the alarm
you must depress the RED button marked RESET. This will silence the alarm and reset
the instrument to begin another test. This RESET button may also be used as a safety
button to quickly ABORT a test and cut off the HIGH CURRENT.
When HIGH CURRENT is present a green TEST indicator located to the
left of the CURRENT receptacle will flash until the HIGH CURRENT is
OFF. If the device under test PASSED, then a single beep will sound and
the GREEN PASS indicator will light. In the case of a FAIL condition, the RED FAIL
indicator will light and the audible alarm will sound. The instrument will provide a
memory of the test condition results on the LCD display that will remain until the next test
is initiated. Depressing the reset button will reset the instrument alarm while keeping the
last test results on the display. Depressing the reset button a second time will clear the
display.