2.
In the Re-injection dialog box, select options, then click OK.
Note:
Sample data files for each re-injection can be saved in a separate folder in
the results group folder if specified in the results group.
If you select a protocol other than the original, the software:
• Creates a copy of the assay specified for the re-injected well (Original_Assay-1).
• Adds the new or modified instrument protocol to Original_Assay-1.
• Assigns Original_Assay-1 to the re-injected well only.
• Saves the plate (the software does not save the copy of the assay to the library).
If the Injection Number attribute is selected for display in the plate view, the number
of the original injection and the re-injection are shown.
1
2
1
Sample 1 selected for re-injection
2
Re-injection number listed for all samples in the re-injection
If you select a
protocol other
than the original
How re-injections
are displayed in
the plate view
Chapter 3
Set up and run
Check sequence or sample quality and re-inject samples
3
70
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1