4.
Click a row in the flag table, then click the Sample tab in Instrument Run Views
to display the associated data in the Sample view.
A re-injection physically re-injects all samples in the capillary array. You can specify
whether to collect data for all or only selected samples in the array.
By default, you can specify a re-injection before the run completes. To allow re-
injections after a run is complete, set the Pause After Last Injection preference (see
“User preferences“ on page 37).
1.
Select the injections or wells to re-inject:
Note:
Re-inject is dimmed if you select an injection that contains more than
one results group, or if you select flags in the flags table that correspond to
samples with different results groups. To enable Re-inject, select samples that
specify the same results group.
To collect data for all wells in an injection
1. Select the injection in the injection
list.
2. Click Re-inject.
To collect data for only specific wells
Note:
You can also re-inject specific
samples in Review Results.
1. Select the injection.
2. Select in the array view the capillary
that corresponds to the well or
sample of interest (see “Array
3. Click Re-inject.
To collect data for only samples that
contain flags
1. Select the samples in the flag table
(see “Check sequence or sample
2. Click Re-inject.
Note:
If you are running an HID plate, see “Re-inject HID allelic ladder
Re
‑
inject samples
from the
Monitor
Run
screen
Chapter 3
Set up and run
Check sequence or sample quality and re-inject samples
3
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
69