5.
If the data for the required number of capillaries do not meet the criteria above
(at least 7 capillaries for 8-capillary instruments, at least 22 capillaries for
24-capillary instruments):
a.
(Optional) If you want to generate a report for the failed calibration, click
View Report before you click Reject Results. To save the report
electronically, select CutePDF as the printer.
b.
Click Reject Results. For troubleshooting information, see “Sequencing
install standard troubleshooting“ on page 274.
IMPORTANT!
If you reject results, the spectral calibration is not saved.
Select History View, then select an install standard to view the associated install
check information.
Example
sequencing install
check results
View previously
run install
standards
Chapter 6
Run calibrations and install checks
Run a Sequencing install check
6
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
129