QC protocols library (primary analysis—HID)
A QC protocol is the required primary analysis protocol for HID applications. A QC
protocol defines peak detection, sizing, and quality values.
When you create an HID assay, you add a QC protocol to the assay. If you add this
item from the library, a copy of the item is added to the assay, and can be modified
independently from the original item stored in the library. For information on how
changes are tracked if auditing is enabled, see “Audit action“ on page 205.
If factory-provided QC protocols do not suit your needs, you can create new QC
protocols:
1.
Access the QC Protocols library.
2.
Click
Create.
3.
In the Analysis Settings tab of the Create New QC Protocol dialog box
(Figure 27), specify settings (“QC protocol—Analysis settings“ on page 187).
4.
Click QC Settings. In the QC Settings tab of the Create New QC Protocol dialog
box (Figure 28), specify settings (“QC protocol—QC settings“ on page 190).
QC protocol
overview
Create a new QC
protocol
Chapter 7
Manage library resources
QC protocols library (primary analysis—HID)
7
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
185