11.
With the tab facing you and the RFID tag to the right, install the CBC on the
autosampler. When properly installed, the CBC tabs will click as you snap them
into place on the autosampler.
A
B
12.
Click the Tray button to retract the autosampler, then close the instrument door
to initialize.
13.
In the Dashboard, click Refresh, then check the Quick View section for updated
status.
Replenish, change, flush, and store polymer
IMPORTANT!
Note the following:
·
Wear appropriate protection, including gloves, laboratory goggles, and coat
whenever you work with the fluids used on this instrument, or parts that may
come into contact with these fluids.
·
Use only the parts listed in Appendix D, “Catalog numbers“.
·
To minimize background fluorescence, use clean, powder-free, silicone-free latex
gloves whenever you handle the pump assembly or any item in the polymer path.
• Do not reuse a polymer pouch that has been installed on another type of
instrument. For example, if you remove a partially used polymer pouch from an
8-capillary instrument, do not reuse that polymer on a 24-capillary instrument.
• If you remove a polymer pouch for storage (2–8°C), place a pouch cap onto the
pouch, then place an empty pouch (or conditioning reagent) on the connector to
prevent desiccation of any residual polymer on the connector. Follow the
instructions in the wizard to ensure proper operation of the pouch and the
instrument.
1.
Check the expiration date on the label to ensure that the polymer is not expired
and will not expire during intended use.
IMPORTANT!
Do not use if the product is expired, if the pouch or label is
damaged, or if the top seal is missing or damaged.
2.
Allow the refrigerated polymer to equilibrate to room temperature (15–30°C)
before use.
3.
In the Dashboard, click Wizards, then click Replenish Polymer (requires 10 to
20 minutes) or Change Polymer Type (requires 60 to 70 minutes).
Precautions for
use
Replenish polymer
or change polymer
type
Chapter 9
Maintain the Instrument
Replenish, change, flush, and store polymer
9
232
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1