Prepare the plate assembly on a clean, level surface. Wear gloves when handling
septa. Do not heat plates that are sealed with septa.
96-well plate assembly
IMPORTANT!
Use the correct plate base for standard plates. Using the wrong plate
base may affect performance. See Appendix D, “Catalog numbers“ for plate assembly
specifications and catalog numbers.
1
2
3
1
Plate retainer
2
Plate with septa strip
3
Plate base
1.
Align the holes in the septa with the wells of the plate (general purpose supply),
then press down firmly on the septa until the septa lies flat on the plate.
2.
Place the plate into the plate base.
3.
Snap the plate retainer (cover) onto the plate, septa, and plate base.
4.
Verify that the holes of the plate retainer and the septa are aligned. If holes are
not aligned, take it apart, then re-assemble.
IMPORTANT!
The array tips will be damaged if the plate retainer and septa
holes do not align correctly.
5.
If the contents of any well contain bubbles or are not located at the bottom of the
well, briefly centrifuge the plate, remove the plate from the centrifuge, and verify
that each sample is positioned correctly in the bottom of its well.
Prepare the plate
assembly
Chapter 3
Set up and run
Prepare and load sample plates
3
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
53