If any consumables are expired or if buffer fill level is too low, replenish the
consumables as described in:
• “Replenish polymer or change polymer type“ on page 232
IMPORTANT!
Wear gloves while handling polymer, the capillary array, septa,
ABC, or CBC.
• “Install the anode buffer container (ABC)“ on page 229
• “Install the cathode buffer container (CBC)“ on page 230
• “Fill capillary array with fresh polymer“ on page 234
• “Install or change the capillary array“ on page 235
When you install the CBC buffer septa, press firmly to seat the septa.
IMPORTANT!
Look at the CBC from the side and ensure that there is no gap between
the container and the lip of the septum.
Replenish
consumables
Ensure proper
installation of CBC
septa
Chapter 2
Start the system
Check system status in the Dashboard
2
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
35