384-well plate assembly
IMPORTANT!
Use the correct plate base for 384-well plates. Using the wrong plate
base may affect performance. See Appendix D, “Catalog numbers“ for plate assembly
specifications and catalog numbers.
1.
Place the sample plate into the plate base.
2.
Place the septum on the plate and press down to seat.
3.
If the reagents of any well contain bubbles or are not located at the bottom of the
well, briefly centrifuge the plate, remove the plate from the centrifuge, and verify
that each sample is positioned correctly in the bottom of its well.
1.
Click the Tray button on the front panel to move the autosampler to the front
position, then open the instrument door.
2.
Place the plate in the autosampler with the labels facing you (or the instrument
door) and the notched corner of the plate in the notched corner of the
autosampler.
3.
Close the instrument door to initialize the instrument.
Load the plate in
the instrument
Chapter 3
Set up and run
Prepare and load sample plates
3
56
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1