IMPORTANT!
If you relocate the instrument, we recommend that you have an IQ OQ
performed. Contact Thermo Fisher Scientific to schedule the IQ OQ service.
WARNING!
PHYSICAL INJURY HAZARD. Do not attempt to lift the
instrument or any other heavy objects unless you have received related
training. Incorrect lifting can cause painful and sometimes permanent back
injury. Use proper lifting techniques when lifting or moving the instrument.
Two or three people are required to lift the instrument, depending upon
instrument weight.
1.
Remove the following components from the instrument:
• Any plate assemblies from the autosampler.
• CBC from the autosampler.
• Capillary array: Click Shutdown the Instrument in the Maintenance
Wizards. See “Shutdown the instrument“ on page 240.
• Anode buffer reservoir.
2.
Switch off the circuit breaker on the back of the instrument.
3.
Disconnect the power cord and the Ethernet cable.
IMPORTANT!
While moving the instrument, avoid any shock or vibration.
4.
Move the instrument.
5.
Turn the instrument legs to level the instrument.
To move the instrument corner ...
Turn the leg ...
up
right (clockwise)
down
left (counterclockwise)
6.
Have an IQ OQ performed before using the instrument.
IMPORTANT!
After performing a conditioning wash, ensure that the buffer level
inside the ABC is at or above fill line before proceeding to the next step.
Move and level the
instrument
Chapter 9
Maintain the Instrument
Shutdown move and reactivate the instrument
9
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
241