4.
Click QV Settings. In the QV Settings tab of the Create New Basecalling Protocol
dialog box (Figure 24), specify settings (see “Basecalling protocol—QV
settings“ on page 178).
QV settings are quality value ranges used in the following screens:
• Monitor Run screen—The state of the QV flag:
– If all three values are in the pass range, the QV flag in Monitor Run is
set to (green).
– If any values are in the suspect range, the QV flag in Monitor Run is set
to (yellow).
– If any values are in the fail range, the QV flag in Monitor Run is set to
(red).
• View Sequencing Results
4
Metric Analysis Results table—The
pass/check/fail status for Trace Score Quality, CRL Quality, and QV20+
Quality results.
Chapter 7
Manage library resources
Basecalling protocols library (primary analysis—sequencing)
7
174
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
Содержание 3500