Chapter 4 Operations for Each Function
4-128
■Total Power display
Figure 4.10.2-6 Total Power Display
The total power of the channels that are set to ON (excluding OCNS)
is displayed in this field. If the total power displayed here is out of
±0.01 dB when OCNS is OFF, the warning message shown below is
displayed during total power calculation, prompting to normalize the
total power.
Figure 4.10.2-7 Warning Display
■Normalize Power button
Figure 4.10.2-8 Normalize Power Button
This button is enabled when OCNS is OFF. When this button is
clicked, the value displayed in Total Power is subtracted from the
power of each channel that is set to ON. As a result, the total power is
normalized to 0 dB while retaining the power ratio between channels.
■Status display
The status display on the bottom of the screen displays operation
statuses such as a warning occurred during setting.